Backtest: would it have seen it coming?
For each public FDA action, the detector is replayed weekly for the 12 weeks before it, on only the reports FDA had received by each date. A flag counts only if it is on one of the three event months before that date, so a spike from an earlier issue is not credited to a later one.
- May 5, 2025iOS app 2.7.0 correction for stale glucose values (Class II, Z-1906-2025)App or display software. A bug could leave the app showing an old glucose value without updating it, so alerts might not trigger. Firm initiated May 5, 2025.missedNot flagged before the event. Strongest sign in the window: Feb 2025 at 1.1x its baseline share (724 reports), short of the flag bar.
- May 12, 2025Receiver speaker recall (Class I, Z-1944-2025)Alert not received. Defective foam or an assembly error may stop the receiver speaker from sounding low and high alerts. Firm initiated May 12, 2025.missedNot flagged before the event. Strongest sign in the window: Dec 2024 at 1.3x its baseline share (401 reports), short of the flag bar.
- Jul 24, 2025G7 app correction for a missing "Sensor Failed" alert (Class I, Z-2447-2025)Alert not received. When the transmitter reported a failure, the app ended the session without a "Sensor Failed" alert, so glucose readings stopped without warning. Firm initiated July 24, 2025.14 days earlyFlagged 14 days before (as of Jul 10, 2025): Jun 2025 at 2.0x its baseline share.
- Apr 14, 2026iOS app correction for delayed glucose values and alerts (Z-2163-2026)Alert not received. A software defect could delay glucose values and alerts while the app ran in the background. Firm initiated April 14, 2026.missedNot flagged before the event. Strongest sign in the window: Feb 2026 at 1.4x its baseline share (1,166 reports), short of the flag bar.
Public reports reach FDA up to 30 days after the manufacturer hears of an event, and recent months fill in slowly. A manufacturer's own complaint feed sees the same reports weeks earlier, which is where this detector belongs. A flag shows that a complaint type rose, not what caused it, and lead time is measured against the public start date of each action. False alarms are not measured here: the replay scans only each action's own mode (see the flag history below).
Every month flagged on today's data
Unlike the backtest, this uses every report received since, so it shows what the detector sees in hindsight. The inserter and applicator rise in early 2025 matches what Dexcom described on its October 2025 earnings call.
| Event month | Mode | Reports | Expected | Ratio |
|---|---|---|---|---|
| Jan 2025 | Sensor wire, needle or applicator | 4,628 | 2,994 | 1.5x |
| Feb 2025 | Sensor wire, needle or applicator | 6,935 | 3,719 | 1.9x |
| Mar 2025 | Sensor wire, needle or applicator | 9,415 | 5,257 | 1.8x |
| Apr 2025 | Sensor wire, needle or applicator | 10,320 | 6,619 | 1.6x |
| Jun 2025 | Alert not received | 1,491 | 616 | 2.4x |
| Jul 2025 | Alert not received | 5,101 | 875 | 5.8x |
| Aug 2025 | App or display software | 3,292 | 1,201 | 2.7x |
| Sep 2025 | App or display software | 1,968 | 1,161 | 1.7x |
| Nov 2025 | Signal loss or pairing | 5,410 | 3,304 | 1.6x |
| Dec 2025 | Signal loss or pairing | 5,585 | 3,462 | 1.6x |
| Dec 2025 | Skin reaction or infection | 56 | 33 | 1.7x |
| Aug 2026 | Skin reaction or infection | 66 | 43 | 1.5x |
Tagging eval
76 real reports, each labeled by hand with one failure mode, scored with golive-gate. Gate: at least 90% right overall and 80% on free-text reports, and no safety report (alerts, inaccurate readings, skin reactions) filed under another mode. Recorded run: claude-haiku-5-5 on Nov 23, 2026.
| Tagger | Gate | Right | Free text | Safety misses | To a person | Cost | Latency | |
|---|---|---|---|---|---|---|---|---|
| Rules alone | NO-GO | 72 of 76 | 22 of 25 | 1 | 0 | free | none | every case |
| Claude, raw answer | GO | 76 of 76 | 25 of 25 | 0 | 0 | $0.181 per 1,000 | 1.48 s | every case |
| Claude with guardrails | GO | 73 of 76 | 23 of 25 | 0 | 3 | $0.181 per 1,000 | 1.48 s | every case |
Read the rules row with care: the free-text reports were found by searching for known complaint wording, so the rules do better here than on wording they have never seen (see below). The guardrails send some answers the model had right to a person: that is the price of never filing a safety report under the wrong mode. The guardrails were fixed after an earlier run on this same set, so treat the guardrail row as tuned, not held out. The model's answers vary a little between runs (75 and 76 of 76 raw across the runs so far), which is why CI re-scores the recorded answers on every change.
What the rules miss
Each run, Claude reads recent reports the phrase rules cannot place. Aug 2026: 200 reports sampled, 34 not placed by the rules. Claude, after guardrails: Sensor failed early 13, Other 12, Signal loss or pairing 5, Sent to a person 2, Receiver hardware 1, Sensor wire, needle or applicator 1. New wording that keeps showing up becomes a new phrase, after a person reviews it.
Phrases added after review
- Nov 23, 2026:
pairing failure occurredadded to Signal loss or pairing. New allegation wording, first common in February 2026 events (166 reports in March 2026 alone). In the run before this phrase was added, Claude placed 10 of the 41 unplaced reports in an August 2026 sample under signal loss, five of them with this wording. - Nov 23, 2026:
poor patch adhesion occurredadded to Adhesion. New allegation wording since March 2026, 14 to 26 reports a month.
How it works
- Count. Each failure mode is a list of phrases (the manufacturer's allegation wording and consumers' own words). openFDA counts every matching report by event month, server-side.
- Detect. A mode is flagged when its share of reports is at least 1.5x its previous six months, with at least 30 reports and a Poisson p below 0.001. Months wait until half their usual reports have arrived.
- Tag. Claude reads narratives the phrases cannot place and picks a mode with a quote as evidence. Guardrails: the mode must exist, the quote must be in the text, low confidence goes to a person, and when the manufacturer's allegation sentence names a known mode, that mode wins.
- Gate. The tagger only ships if it passes the eval gate in CI.
Limits: openFDA narratives are blank for many of the latest reports, which can hide a rise but not create one. Report counts are not failure rates. This page reads public data only and says nothing about causes.